Abstract
The article presents an analysis of the existing approaches in establishing a correlation between a calculation method and
the in vitro value of effective orifice area (EOA) of heart valve prostheses. This is particularly important for newly developed
or modified valve designs and in indexing EOA values to the patient's body surface area (BSA) when selecting the type and
the size of prosthesis for an individual. The findings of the EOA calculations by a variety of techniques are reported.
Following are the methods and evaluation of EOA values used among other parameters for the minimum performance
requirements to be brought into Technical Standards for heart valve prostheses when such Standards are intended to be
revised or amended.
References
Агафонов А. В., Литвинов М. М. Интегральная оценка гидродинамических свойств протезов аортальных клапанов // Механика композитных материалов. 1979. № 3. С. 524-527.
Бокерия Л. А., Николаев Д. А., Фадеев А. А. Исследование протезов клапанов сердца в сходящемся закрученном стационарном потоке // Протезы клапанов сердца МедИнж в хирургии клапанных пороков сердца: сб. трудов. Ч. 2. М.: НЦССХ им. А. Н. Бакулева РАМН, 2009. С. 162-167.
Бокерия Л. А., Фадеев А. А., Николаев Д. А. и др. Развитие технических требований к протезам клапанов сердца // Бюллетень НЦССХ им. А. Н. Бакулева РАМН. 2009. Т. 10, № 6. С. 64-70.
Gabbay S., McQueen D., Yellin E., Frater R. In vitro hydrodynamic comparison of mitral valve bioprostheses // Circulation. 1979. Vol. 60, № 2. P. 62-70.
House C. M., Nelson W. B., Kroshus T. J. et al. Manufacturer-provided effective orifice area index charts and the prevention of prosthesis- patient mismatch // J. Heart Valve Dis. 2012. Vol. 21. P. 107-111.
Marquez S., Hon R., Yoganathan A. Comparative hydrodynamic evaluation of bioprosthetic heart valves // J. Heart Valve Dis. 2001. Vol. 10. P. 802-811.
Pibarot P., Dumesnil J. G. Prosthesis-patient mismatch: definition, clinical impact, and prevention // Heart. 2006. Vol. 92. P. 1022-1029.
Yoganathan A., Chaux A., Gray R. et al. 1984. Bileaflet, tilting disc and porcine aortic valve substitutes: in vitro hydrodynamic characteristics // JACC. 1984. Vol. 3, № 2. P. 313-320.
Yoganathan A., Travis B. Fluid dynamics of prosthetic valves // The practice of clinical echocardiography. Philadelphia: W. B. Saunders Company, 2002.